|
| Home | NSE | Instruments | NSEpol418 | Contacts | SiteMap |
|
Polarizer and analyzer efficiencies on NG-5 NSE
The polarizer on the NG-5 NSE is a Mezei cavity: a 1500 mm section of
58Ni-equivalent guide that contains a polarizing supermirror
“V” 1253 mm long. The polarizing “V” is made from 1 mm thick Si
wafers that are coated on one side with
3 The polarizing elements have been installed by Cilas so that the supermirror is on the outside of the “V” (on the side closest to the guide section walls), with no overlap of the two arms at the tip of the “V”. As a result, the center of the beam is only slightly polarized, as can be seen in Figure 1.
We removed this gap by placing a piece of B-loaded Al at the tip of the “V”, thereby removing the worst of the effect, with only a slight decrease in the flipping ratio toward the center of the guide, as shown in Figure 2.
In testing a piece of ILL Co/Ti polarizer on the spectrometer (located in place of the analyzer), I found that the flipping ratio of an elastically scattered beam was much larger (R =12-13) than that of the direct beam (R = 9). I started to do some Monte Carlo simulations of the polarizing cavity to see if I could reproduce this behavior by mispositioning the mask. I performed the simulations with two different image planes. In the first, I placed the image plane at the sample position. This, after integrating over the sample area, corresponds to simulating the behavior of the scattered beam, since in the actual measurement, the sample coherently scatters into a single detector. The direct beam measurement (with no sample) is modeled when the image plane is placed at the analyzer. In our tests with the ILL supermirror, only the polarization at the center of the detector is measured, since the supermirror subtends a width of only about 2mm. The original simulations of the Mezei cavity had the polarizing supermirror on the inside of the “V”, but simulations with the supermirror on the outside of the “V” and the neutrons first incident on the Si, showed that there were many low-divergence neutrons with the wrong spin state passing through the polarizing cavity. These are apparently due to a dip in the + (parallel) spin state reflectivity at around the Si critical edge when the neutrons impinge on the supermirror through the Si substrate (see Figure 3).
If we assume that the flipping ratio is 24 (P = 0.92) for the ILL test piece (as Ian Anderson reported) we then expect to observe flipping ratios of 12.1 (P = 0.847) for the scattered beam and 9.2 (P = 0.803) for the direct beam with the test piece, in good agreement with experiment. We can then use the observed flipping ratio with the Fe/Si analyzer (in a too-small field) to determine that the analyzer polarization is 0.77, giving our observed flipping ratio of 6 in the scattered beam and 5.4 in the direct beam. If the Mezei cavity is flipped to remove this problem, we should get a cavity polarization of 0.97 and so a final polarization of 0.747, increasing the flipping ratio to 6.9. With the new analyzer polarization of 0.92, R = 17.6.
Download a postscript version of this report. | |||||||||||||||||||||||||||||||||||||||||||||